Binder CB 60 Manual do Utilizador Página 114

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 113
CB (E6) 09/2013 Page 114/144
This symbol on the controller display indicates that the hot-air sterilization was aborted. The
interior may still be hot. Do not touch the glass door and inner surfaces!
Fixed value
08.03.2013 05:05:06
Initial view after aborting the hot-air sterilization
(sample values).
For further information, press the “Info” button.
Setpoint
Actual value
Temperature [°C]
37.0
60.2
CO2 [Vol.-%]
5.0
5.0
Humidity [%rH]
85.5
User
Info
Menu
Sterilization information
08.03.2013 05:05:06
Information menu “Sterilization information”.
Confirm the information with “Ok”.
Sterilization information
STERILIZATION ABORTED!
Close Ok Home
If the sterilization procedure is aborted prematurely, whether effective sterilization has occurred depends
on the time that has elapsed:
Aborting sterilization after less than 4 hours: Prevents effective sterilization.
Aborting sterilization after more than 4 hours: The unit is definitely in the cooling-down phase, meaning
that the necessary duration for the proper sterilization phase has occurred.
Aborting sterilization during the cooling-down phase (after more than 4 hours)
The duration of the entire sterilization is approx. 10 hours. If you want to shorten the sterilization
procedure in order to save time, you can abort it during the cooling-down phase, i.e. no sooner than after
4 hours. At this point, the inner temperature is still approx. 140 °C / 284°F.
CAUTION
Glass door and inner chamber become hot during sterilization.
Danger of burning.
Do NOT touch the glass door and inner surfaces for approx. 4 hours after aborting
sterilization.
Vista de página 113
1 2 ... 109 110 111 112 113 114 115 116 117 118 119 ... 144 145

Comentários a estes Manuais

Sem comentários